POWER-WIN TECHNOLOGY CORP.
|
Item Number.:
|
PW-IM040/065B2-1YXXXZ |
Product Name:
|
40/65W ITE & Medical power supply, 2x 4, DOE level VI and EU ErP tier 2 Efficiency Compliant |
Description:
|
*Both ITE & Medical Approvals.
*Medical Body Floating ( BF ) rated, Class I Power Supply
*Medical applications Protection: Means of patient protection ( MOPP )
*Altitude during operation: ITE up to 5000m ,Medical Below 3000m.
*Less than 75mW no-load power consumption
*Low leakage current
*3 year warranty
|
Specification:
|
Electrical specifications
- Input range : 90 - 264VAC
- Frequency : 47 - 63Hz
- Inrush current : < 45A peak @115VAC; < 90A peak @230VAC cold start @25°C
- Input current ( rms ) : 2A @115VAC; 1A @230VAC max.
- Efficiency : > 87% typical @full load, 115VAC ( 5V series excluded )
- Touch current < 100uA @264VAC
- Maximum output power ( Po ) : 65 Watts convection cooling
- Hold-up time : > 10ms typical @full load, 115VAC
- Short circuit protection : Auto-recovery
- Over load protection : 110% to 180% maximum rating, Auto-recovery
- Over voltage protection : Latching type. AC Recycle
Safety standards
- UL60601-1 3.1rd Edition
- UL/c-UL UL60950-1
- UL/c-UL UL62368-1
- EN60601-1 3.1rd Edition
- TUV EN60950-1
- TUV EN62368-1
- IEC EN60601-1 3.1rd Edition
- CB IEC 60950-1
- CB IEC 62368-1
EMC standards
- EN60601-1-2
- EN55024
- EN 55011 Class B
- EN55032 Class B
- FCC Part 15 Class B
- FCC Part 18 Class B
- CE
Environmental
- Operating temperature :-20 to +70°C ( Derate linearly 2.5% per °C from 51 to 70°C, Derate linearly 1.0% per °C from 0 to -20°C )
- Storage temperature: -10°C to +85°C
- Humidity: Non-condensing 10% to 90%
- MTBF : > 150,000 hours @full load and 25°C ambient temperature
RoHS compliant
- Dimension : L101.6 × W50.8 × H27.5 mm ( 4" x 2" x 1.08" )
- Weight : 0.136 kgs. ( 0.30 lbs. )
DC OUTPUT & FEATURES
1. In/Output connector options: Z=J ( JST VH Type Connector or Equ. ), Z=M ( TKP P8800I type) 1. Ripple and noise are measured at oscilloscope 20MHz bandwidth by a 10uF electrolytic capacitor and a 0.1uF ceramic capacitor in parallel at output connector. 2. The ripple and noise will be higher while load under 30% of rate load during Burst Mode operation. The Burst Mode operation is to meet the latest efficiency regulations. 3. The output voltage is set by manufacturer
|
|
|
|
|
 |
Send Message
(Your message will be sent directly)
|
|
|
|
|
|