Trade Asia Homepage
POWER-WIN TECHNOLOGY CORP.
40/65W ITE & Medical power supply, 2x 4, DOE level VI and EU ErP tier 2 Efficiency Compliant





>> Company profile
>> View more products
Item Number.:
PW-IM040/065B2-1YXXXZ
Product Name:
40/65W ITE & Medical power supply, 2x 4, DOE level VI and EU ErP tier 2 Efficiency Compliant
Description:
*Both ITE & Medical Approvals.
*Medical Body Floating ( BF ) rated, Class I Power Supply
*Medical applications Protection: Means of patient protection ( MOPP )
*Altitude during operation: ITE up to 5000m ,Medical Below 3000m.
*Less than 75mW no-load power consumption
*Low leakage current
*3 year warranty
Specification:
Electrical specifications
  • Input range : 90 - 264VAC
  • Frequency : 47 - 63Hz
  • Inrush current : < 45A peak @115VAC; < 90A peak @230VAC cold start @25°C
  • Input current ( rms ) : 2A @115VAC; 1A @230VAC max.
  • Efficiency : > 87% typical @full load, 115VAC ( 5V series excluded )
  • Touch current < 100uA @264VAC
  • Maximum output power ( Po ) : 65 Watts convection cooling
  • Hold-up time : > 10ms typical @full load, 115VAC
  • Short circuit protection : Auto-recovery
  • Over load protection : 110% to 180% maximum rating, Auto-recovery
  • Over voltage protection : Latching type. AC Recycle
Safety standards
  • UL60601-1 3.1rd Edition
  • UL/c-UL UL60950-1
  • UL/c-UL UL62368-1
  • EN60601-1 3.1rd Edition
  • TUV EN60950-1
  • TUV EN62368-1
  • IEC EN60601-1 3.1rd Edition
  • CB IEC 60950-1
  • CB IEC 62368-1
EMC standards
  • EN60601-1-2
  • EN55024
  • EN 55011 Class B
  • EN55032 Class B
  • FCC Part 15 Class B
  • FCC Part 18 Class B
  • CE
Environmental
  • Operating temperature :-20 to +70°C ( Derate linearly 2.5% per °C from 51 to 70°C, Derate linearly 1.0% per °C from 0 to -20°C )
  • Storage temperature: -10°C to +85°C
  • Humidity: Non-condensing 10% to 90%
  • MTBF : > 150,000 hours @full load and 25°C ambient temperature
RoHS compliant
  • Dimension : L101.6 × W50.8 × H27.5 mm ( 4" x 2" x 1.08" )
  • Weight : 0.136 kgs. ( 0.30 lbs. )
DC OUTPUT & FEATURES
Model Output Voltage Minimun Load Maximun
Load
Po Average efficiency Output
Regulation
Ripple & Noise (Vp-p)
Full Load < 30% Load
PW-IM040B2-1Y050Z +5V 0A 8.00A 40W > 86% ±5% 100mV 200mV
PW-IM065B2-1Y120Z +12V 0A 5.42A 65W > 89% ±3% 120mV 200mV
PW-IM065B2-1Y150Z +15V 0A 4.34A 65W > 89% ±3% 120mV 300mV
PW-IM065B2-1Y190Z +19V 0A 3.43A 65W > 89% ±3% 120mV 300mV
PW-IM065B2-1Y240Z +24V 0A 2.71A 65W > 89% ±3% 200mV 300mV
PW-IM065B2-1Y280Z +28V 0A 2.33A 65W > 89% ±3% 200mV 350mV
PW-IM065B2-1Y480Z +48V 0A 1.36A 65W > 89% ±2% 200mV 480mV
PW-IM065B2-1Y540Z +54V 0A 1.21A 65W > 89% ±2% 200mV 480mV
1. In/Output connector options: Z=J ( JST VH Type Connector or Equ. ), Z=M ( TKP P8800I type ) 1. Ripple and noise are measured at oscilloscope 20MHz bandwidth by a 10uF electrolytic capacitor and a 0.1uF ceramic capacitor in parallel at output connector. 2. The ripple and noise will be higher while load under 30% of rate load during Burst Mode operation. The Burst Mode operation is to meet the latest efficiency regulations. 3. The output voltage is set by manufacturer
 
 
Send Message    (Your message will be sent directly)  
* Required information     
* Subject :
* Message :
 Please enter only English message between 20 and 900 characters
Contact Information
* Contact Person :
Please enter First and Last Name
* Email :
* Company Name :
* Country/Region :
* Industry :
Telephone:
Web Site :
* Please enter your check code :

More Exposure,
More Buyers Match,
Unlimited Number of Products.
Apply now !
Wonpro Co., Ltd.

Copyright (c) 2019 TradeAsia. All Rights Reserved.