POWER-WIN TECHNOLOGY CORP.
|
Item Number.:
|
PW-IM065B-1YXXX |
Product Name:
|
40/65W ITE & Medical open frame power supply, 2x3, DOE level VI and ErP Tier 1 efficiency compliant |
Description:
|
Power-Win is a professional power supply manufacturer located in Taipei, Taiwan. Starting with PC power supplies business in 1989, Power-Win extends its product lines into Desktop Power Supply, open frame power supplies, DC / DC ATX converters, and medical power supplies. With over 20 year experience in design, research, and manufacturing, Power-Win not only develops its own products to lead market trends and also works with customers for OEM / ODM projects. To meet latest energy saving demand and green power trend, we introduce more efficiency level V adapters and also open frame units to fulfill customer’s needs.
Etc. PW-M065B-1Yxxx (Medical Power Supply)
*Both Medical 60601-1 3.1rd Edition MOPP and ITE approved
*Compact, High Power density, 40W/65W in 2" x 3" footprint
*Meet DOE level VI 88% & ErP Tier 1 requirements.(Except +5V mode 86%)
*No load power consumption < 0.21W@230VAC (Except +5V mode ,< 0.1W@230VAC max)
*Ultra low Earth leakage current < 189uA @264VAC
*2 year warranty
|
Features:
|
*Dimension : L76.2 × W50.8 × H28.7 mm ( 3" × 2" × 1.13" )
*Weight : 0.12 kgs. ( 0.26 lbs. )
|
Specification:
|
Electrical specifications
- Input range : 90 - 264VAC
- Frequency : 47 - 63Hz
- Inrush current : < 65A peak @115VAC; < 130A peak @230VAC cold start @25°C
- Input current (rms) : 1A @115VAC; .0.6A @230VAC max.
- Efficiency : > 87% typical @full load, 115VAC(Except +5V mode ,85% typical @full loadl)
- Maximum output power (Po) : 65 Watts convection cooling((Except +5V mode ,40 Watts )
- Hold-up time : > 10ms typical @ load 75% , 115VAC
- Short circuit protection : Auto-recovery
- Over load protection : 105% to 170% maximum rating, Auto-recovery
- Over voltage protection : Shutdown and latch off, AC recycle
Environmental
- Operating temperature : -20 to +70°C (Derate linearly 2.5% per °C from 41 to 70°C)
- Storage temperature: -20°C to +85°C
- Humidity: Non-condensing 0% to 90%
- MTBF : > 100,000 hours @full load and 25°C ambient temperature
SAFETY STANDARDS
Medical
- 60601-1 3.1rd Edition
- CSA-C22.2 No.60601-1 3.1rd Edition
- EN60601-1 3.1rd Edition
- IEC EN60601-1 3.1rd Edition
ITE
- UL UL62368-1
- CSA-C22.2 NO.62368-1
- EN62368-1
- CB IEC62368-1
EMC STANDARDS
Medical
- IEC60601-1-2 Ed4:2014
- EN60601-1-2 Ed4:2015
- FCC Part 18 Class B
- EN55011 Class B
- CE
ITE
- EN55032 Class B
- CISPR22 Class B
- FCC Part 15 Class B
- CE
Model no. |
Output
voltage |
Minimum
load |
Maximum
load convection |
Po |
O/P
Regulation |
Ripple & Noise
(Vp-p) |
Efficiency Level |
PW-IM040B-1Y050 |
+5V |
0A |
8A |
40W |
±5% |
150mV |
VI |
PW-IM065B-1Y120 |
+12V |
0A |
5.416A |
65W |
±3% |
240mV |
VI |
PW-IM065B-1Y150 |
+15V |
0A |
4.333A |
65W |
±3% |
300mV |
VI |
PW-IM065B-1Y190 |
+19V |
0A |
3.421A |
65W |
±3% |
300mV |
VI |
PW-IM065B-1Y200 |
+20V |
0A |
3.25A |
65W |
±3% |
300mV |
VI |
PW-IM065B-1Y240 |
+24V |
0A |
2.708A |
65W |
±3% |
300mV |
VI |
PW-IM065B-1Y280 |
+28V |
0A |
2.321A |
65W |
±3% |
300mV |
VI |
PW-IM065B-1Y480 |
+48V |
0A |
1.354A |
65W |
±3% |
300mV |
VI |
PW-IM065B-1Y560 |
+56V |
0A |
1.161A |
65W |
±3% |
300mV |
VI |
Note:
1. At 25°C including initial tolerance, line voltage, load currents and output voltages adjusted to factory settings.
2. Ripple and noise are measured at oscilloscope 20MHz bandwidth by a 100uF electrolytic capacitor and a
0.1uF ceramic capacitor in parallel at output connector.
3. DC OUTPUT & FEATURES Operating temperature is 0 to +40°C.
4. -20 to 0°C output characteristics Output regulation are ±10%.
|
|
|
|
|
 |
Send Message
(Your message will be sent directly)
|
|
|
|
|
|